Patient Monitor (ACCU1200 plus) Display TFT 12''- 7 functions- For Adult, Pediatric, Neonate- Trend 240 hr - Including built in rechargeable battery & thermal printer (Black) - by AKAIMED
Display: Size: 12.1 inch color TFT Resolution: 800 x 600 pixels
Indicators: Categorized alarms (3 priority levels) Visible and audible alarm Battery Power Indicator
Battery: 2.5 hours rechargeable Li-ion battery
Trend Time: 240 hr Recorder: Thermal array (Up to 3 channels)
Data Transfer: USB OTG port .SD cart slot and USB port ACCU-1200
Parameters: ACCU-1200 Plus Parameters:
Standard parameters:ECG, RESP, NIBP, SPO2, TEMP, CO2, IBP ECG: Lead Mode 3-Lead: I, II, III 5-Lead: I, II, III, aVR, aVL, aVF, V Waveform 3-Lead:
1-channel waveform 5-Lead: 2-channel waveform, max. 7 waveforms; Waveform Speed 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
Heart rate: Measurement Range ADU: 15 bpm ~ 300 bpm
PED: 15 bpm ~ 350 bpm Accuracy ±1% or ±1 bpm, whichever is greater Resolution 1 bpm
ST segment analysis: Measurement Range: -2.0 mV ~ +2.0 mV Arrhythmia analysis 16 different arrhythmia analysis classification (applicable to adult and pediatric)
SPO2: Measurement Range 0 ~ 100 % Accuracy ±2 % (70%~100% SpO2) Undefined (0~69% SpO2) Pulse Rate Range 20bpm ~ 300bpm Pulse Rate Accuracy ±3bpm
Respiration: Measurement method: Trans-thoracic impedance RR measuring range: Adult 0 to 120 rpm Pediatric 0 to 150 rpm
Apnea Alarm: Yes NIBP: Measurement Method: Oscillometric Operation mode: Manual/Automatic/Continuous Cuff pressure measuring range: 0 mmHg ~ 300 mmHg Measurement Range: Adult mode SYS: 40 mmHg ~ 270 mmHg DIA: 10 mmHg ~ 215 mmHg MAP: 20 mmHg ~ 235 mmHg Pediatric mode SYS: 40 mmHg ~ 200 mmHg DIA: 10 mmHg ~ 150 mmHg MAP: 20 mmHg ~ 165 mmHg TEMP: Measurement method: Thermal resistance Channel: 2 Measuring Range: 0.0 °C ~ 50.0 °C Resolution: 0.1°C Accuracy (Without sensor): +/-0.1°C CO2: (Standard only Accu1200 Plus) Measurement method Infra-red Absorption Technique CO2 Measuring Range 0~ 20% (0 mmHg ~ 150 mmHg) CO2 Measuring Accuracy < 5.0% CO2: ± 2 mmHg > 5.0% CO2: < 6% of the readings