MED SCAN R FOR VETERINARY

TECHNICAL SPECIFICATION.

Type High Frequency 50kHz X-ray Generator
Power Output 6 KW
KV RANGE 40 to 120 KVP
mA Output
FLURO Normal Mode 0.1 to 5mA
HD/Boost
Flouro mode up to 10mA
Radiography Up to 120mA
mAS Range 1-200mAS

1. CONTROL PANEL
A very compact, soft touch control panel(APR) with L.C.D
display on which KV, fluoro mA, radiography mAs, fluoro
time, I.I Zoom, error interlocks for KV, filament and thermal
overload are displayed. Console panel has following functions
and indications.
o Machine ON/OFF switch.
o I.I magnification (I.I field) selection switch.
o Fluoro and Radio mode selection.
o Fluoroscopy timer (Five minute cumulative timer with buzzer that activates after the completion of 300 seconds of exposure to reinitiate the exposure reset switch is provided).
o ABS (Automatic brightness stabilization) selection for hand free operation- also known as ADR
o KV and mAs increase and decrease switches.
o X-Ray ON switch with indicator.
o Switches for up/down movement of "C" on both side of panel.
o Radio and Fluoro Exposure ON Switches on Panel
o Emergency OFF switches on the control panel.

2. X-RAY TUBE Anode type
Rotating Anode
Small focal spot:0.3mm (Max. power 5 KW)
Large focal spot:0.6mm (Max. power 17 KW)
Anode heat storage capacity: 300 KHU
The X-Ray Tube Head is thermally protected.

3. COLLIMATOR Iris collimator

4. X-RAY IMAGE INTENSFIER
Nominal entrance field size 23 cm (9")
Output image diameter 25mm
Field size( Three fields are available Normal,Zoom1 and Zoom2)
Normal = 9"
Zoom-1 = 6"
Zoom-2 =4.5"
Resolution 52Lp/cm
 
5. CAMERA High definition camera

6. MONITER
32" High resolution LCD monitor
with split screen display with trolley

7. C-ARM MOBILE STAND
FOCUS TO I.I DISTANCE 940mm
C-DEPTH 600mm
745mm
ROTATIONAL MOVEMENT +/-180 0
MOTORIZED UP/DOWN TRAVEL 430mm
HORIZONTAL TRAVEL 210mm
ARC ORBITAL MOVEMENT +90 to -300
WIG WAG(PANNING) MOVEMENT +/-12.50

8. MEMORY FEATURES
High Resolution image processing software
Main features
o GUI based software panel for easy operation
o Pre-set image parameters setting as per Extremities, Ortho, Uro procedures.
o Unlimited patient folder creation with total image storage up to 500 images
Pre-processing:-
o LIH Operating Mode
o Averaging 1 to 16 frames
o Patient data Entry, Manual & Emergency modes
o Pulse Fluoroscopy & Cine loop recording.
Post processing:-
o Brightness /Contrast Adjustments/Enhancement with control of WW & WL
o Negative image feature
o zoom with pan
o Image Flip UP/DN & L/R
o Image rotation feature
o Quad view
o Image save & Retrieve from & to separate patient images folders
Storage & Connectivity:-
o PC connectivity through LAN port
o Image storage through USB

9. POWER SUPPLY REQUIRMENT
230Volt, 1phase supply of 50/60Hz Hz Frequency with line regulation of +/-10%